Intensive Care  ·  Operating Room  ·  Cath Lab

Disposable Pressure Transducer (DPT)

Invasive Blood Pressure (IBP) Monitoring Set

A sterile, single-use pressure transducer that converts a patient's arterial or central venous pressure into an electrical signal for continuous, real-time waveform monitoring on the bedside monitor — a more precise and immediate picture than an intermittent cuff reading.

Single Use Sterile Latex-Free DEHP-Free Type CF Applied Part
Request Information →
Disposable Pressure Transducer (DPT) kit
9
Connector Options
3
Channel Configurations
±4 mmHg
Measurement Accuracy
96h
Max. Use Duration
What It Is

A more accurate window into blood pressure

The DPT sits between a patient's arterial or central-venous catheter and the bedside monitor. Instead of an intermittent cuff reading, clinicians see a live, continuous pressure waveform — essential for patients whose blood pressure needs to be watched moment to moment.

The kit includes everything needed for setup: a 120 cm pressure tube with a 30 cm extension, a continuous 3 mL/hour flush device to keep the line clear, and a monitor cable connector matched to the hospital's existing patient monitors.

Full DPT fluid pathway
Why This DPT

Built for reliable, everyday ICU use

1

DEHP-Free & Latex-Free

Safer for sensitive and pediatric patient populations, with no plasticizer leaching risk.

2

Closed Blood Sampling

An optional closed sampling reservoir reduces infection risk and avoids wasting the patient's blood.

3

Consistent Build Quality

Every component is produced under the same quality system, from part to finished, sterile kit.

4

Broad Monitor Compatibility

Nine connector options cover Abbott, Edwards, BD, Philips/B.Braun and more — no new hardware needed.

Configurations

One, two or three pressure lines at once

Every kit is single-use, supplied sterile, and includes the same 120 cm pressure tube, 30 cm extension and continuous flush device.

Single Channel DPT
PT-01 Series

Single Channel

Arterial line (ABP) monitoring.

Dual Channel DPT
PT-02 Series

Dual Channel

ABP + CVP simultaneous monitoring.

Triple Channel DPT
PT-03 Series

Triple Channel

Multi-site hemodynamic monitoring.

Blood sampling reservoir on the DPT line
Optional Add-On

Closed, needle-free blood sampling

A version of the kit adds a blood sampling reservoir directly on the pressure line — letting staff draw an arterial sample without disconnecting the system or exposing themselves to a needlestick.

  • Needleless design — no separate flushing device required after sampling.
  • Self-closing reservoir — flushes automatically during infusion to reduce residual blood in the line.
  • Negative-pressure return — conserved blood is pushed back to the patient instead of discarded.
Type I reservoir

Type I

Standard twist-knob design — turn counter-clockwise to draw, clockwise to return.

Draw volume: 5 mLChannels: Single · Dual · Triple
Type II reservoir

Type II

Push-pull plunger, Edwards-style — pull to draw, push to return.

Draw volume: 5 mLChannels: Single
Type III reservoir

Type III

Push-pull plunger, Abbott-style — pull to draw, push to return.

Draw volume: 10 mLChannels: Single
The Kit

The main components

Flush device, chip and zeroing stopcock

Flush Device · Chip · Zeroing Stopcock

Clears air, senses pressure, and references the sensor to atmosphere.

Cable connector

Cable Connector

Plugs into the monitor's pressure module via the interface cable.

Stopcock and blood sampling port

Stopcock & Sampling Port

Directs flow through the line and takes a syringe or vacuum tube.

Blood sampling reservoir

Blood Sampling Reservoir

Draws and returns blood on kits supplied with the reservoir option.

Compatibility

Works with the monitors already in your department

ACE / USB Utah Argon Abbott Edwards BD Philips / B.Braun PVB Mindray
Where It's Used

Clinical Indications

SettingApplication
ICUContinuous arterial or central venous pressure monitoring
Operating RoomIntraoperative hemodynamic monitoring
Cath LabProcedural pressure monitoring
Critical & Vascular CareClosed, needle-free blood sampling without added blood loss
The Numbers

Specifications

ParameterValue
Accuracy±4 mmHg @ 1 Hz
Sensitivity5.0 μV/V/mmHg ±1%
Operating Pressure−50 to +300 mmHg
Flush Rate3 mL/hour, continuous
Storage Condition−18°C to 50°C, 30–80% RH
Max. Use Duration96 hours (replace after)